Comparison of setup accuracy and intrafraction motion using stereotactic frame versus 3-point thermoplastic mask-based immobilization for fractionated cranial image guided radiation therapy
2013
Abstract Purpose Prospectively compare patient setup accuracy and intrafraction motion of a standard 3-point thermoplastic mask with the Gill-Thomas-Cosman relocatable stereotactic frame, during fractionated cranial radiation therapy using the ExacTrac system (Brainlab AG Feldkirchen, Germany) for daily online correction. Methods and Materials The number of fractions with all postcorrection and post-treatment errors Results With daily online correction, patient setup margins can be reduced from 1, 5, and 4 mm in the lateral, longitudinal, and vertical axes for mask setup and from 1-2, 2, and 1 mm, respectively, for frame setup to Conclusions Daily online correction can compensate for less precise immobilization and permits stereotactic margins to be used for standard thermoplastic masks without the need for specialized mask systems.