Wide energy range gamma-ray calibration source
2011
Calibration source with monoenergetic gamma-ray lines in wide energy range designed for gamma-ray detector energetic calibration and testing has been built. Gamma-rays are obtained from thermal neutron capture, which is a suitable and cost efficient way how to provide discrete gamma-ray lines with energies above 3 MeV with reasonable intensity. With appropriate and interchangeable targets the source can generate different gamma-ray spectra with energy up to 10 MeV. We present the data obtained with neutron capture on chlorine, but other elements with high thermal neutron capture cross-section such as chrome, iron, nickel and titanium can be used as well. As neutron source we employ radionuclide sources (252Cf or 241Am-Be) with emission rate about 106 neutrons/s. The emitted fast neutrons are moderated by a moderator made of light materials such as graphite, standard water or heavy water. Performance of the source is demonstrated by calibration spectra measured by HPGe and scintillation detectors (LaBr3, NaI(Tl)).