Accuracy of intraocular lens power calculation for cataract surgery after deep anterior lamellar keratoplasty

2021 
Background The purpose of this study was to assess the accuracy of intraocular lens power (IOL) formulas for cataract surgery after deep anterior lamellar keratoplasty (DALK). Methods This retrospective study included eyes which had previously undergone DALK and underwent standard phacoemulsification with monofocal IOL implantation between January 2012 and January 2021 at Ospedali Privati Forli "Villa Igea" (Forli, Italy). The predicted spherical equivalent (SE) was calculated using the Barrett Universal II, Emmetropia Verifying Optical (EVO), Haigis, Hoffer Q, Hoffer QST, Holladay 1, Holladay II, Kane and SRK/T formulas. Prediction error (PE) was calculated as the actual postoperative SE refraction minus the SE predicted refraction. Results 82 eyes of 82 patients were included. The mean PE was negative using all formulas. Friedman test revealed a statistically significant difference of the median absolute PE (MedAE) among the different IOL formulas (P = 0.005). On the basis of the MedAE, the fomulas were ranked as follows: SRK/T (0.805 D), Kane (0.810 D), EVO (0.845 D), Hoffer QST (0.847 D), Barrett (0.895 D), Holladay 1 (0.915 D), Haigis (1.010 D) and Hoffer Q (1.070 D) formulas. Conclusions All formulas had a tendency towards a myopic refractive surprise in post-DALK eyes. Although the SRK/T, Kane, EVO and Hoffer QST formulas were more accurate, predictability of refractive outcomes was lower than in virgin eyes. This article is protected by copyright. All rights reserved.