Clark electrode

The Clark electrode is an electrode that measures ambient oxygen concentration in a liquid using a catalytic platinum surface according to the net reaction: The Clark electrode is an electrode that measures ambient oxygen concentration in a liquid using a catalytic platinum surface according to the net reaction: It improves on a bare platinum electrode by use of a membrane to reduce fouling and metal plating onto the platinum. Leland Clark (Professor of Chemistry, Antioch College, Yellow Springs, Ohio, and Fels Research Institute, Yellow Springs, Ohio) had developed the first bubble oxygenator for use in cardiac surgery. However, when he came to publish his results, his article was refused by the editor since the oxygen tension in the blood coming out from the device could not be measured. This motivated Clark to develop the oxygen electrode. The electrode, when implanted in vivo, will reduce oxygen and thus required stirring in order to maintain an equilibrium with the environment. Severinghaus improved the design by adding a stirred cuvette in a thermostat. A discrepancy between the measured partial pressure of oxygen (pO2) between blood samples and gaseous mixtures of identical pO2, meant that the modified electrode required calibration; consequently a microtonometer was added to the water thermostat. The electrode compartment is isolated from the reaction chamber by a thin Teflon membrane; the membrane is permeable to molecular oxygen and allows this gas to reach the cathode, where it is electrolytically reduced.